Flowrence® products specifications
![FlowrenceSystem](https://rds.avantium.com/wp-content/uploads/2024/07/FlowrenceSystem.jpg)
Reactor Section
Easy and quick reactor exchange system. Possibility to use quartz reactors at high pressure.
1 block of 4 reactors
HT = High Temperature max. 800°C nominal, limited to 925°C (<0.5°C reactor to reactor deviation)
Temperature Ranges (°C)
100 – 800°C
up 925°C (Option)
50 – 525°C
100 – 800°C
up 925°C (Option)
50 – 550°C
50 – 550°C
Reactor Types
L= Length
OD= Outer Diameter
ID= Inner Diameter
SS= Stainless Steel (< 550⁰C)
Qz= Quartz (< 925⁰C)
L | 300 mm | 561 mm |
OD | 3 mm | 6 mm |
ID SS | 2 / 2.6 mm | 2 / 3 / 4 / 5 mm |
ID Qz | 2 mm | 2 / 4 mm |
300 mm | 561 mm | 561 mm |
3 mm | 3 mm | 6 mm |
2 / 2.6 mm | 2 / 2.6 mm | 2 / 3 / 4 / 5 mm |
2 mm | 2 mm | 2 / 4 mm |
561 mm |
3 mm |
2 / 2.6 mm |
2 mm |
561 mm |
3 mm |
2 / 2.6 mm |
2 mm |
Maximum Catalyst Bed Length
(isothermal zone tolerance ± 1°C)
Note: isothermal length is dependent on the temperature range
300 / 3 HT | 561 / 6 HT |
>120 mm @ 450°C | >200 mm @ 500°C |
>90 mm @ 800°C | >150 mm @ 800°C |
>140 mm @ 925°C |
300 / 3 HT | 561 / 3 MT | 561 / 6 HT |
>120 mm @ 450°C | >310 mm @ 450°C | >200 mm @ 500°C |
>90 mm @ 800°C | >150 mm @ 800°C | |
>140 mm @ 925°C |
561 / 3 MT iRTC |
250°C ±0.5°C 41cm (4reactors) |
350°C±0.5°C 38cm (4reactors) |
550°C±0.5°C 28cm (4reactors) |
3 reactors at 550°C, 1 reactor 350°C: 550°C=27cm 350°C=41cm ±0.5°C |
561 / 3 MT iRTC |
250°C ±0.5°C 41cm (4reactors) |
350°C±0.5°C 38cm (4reactors) |
550°C±0.5°C 28cm (4reactors) |
3 reactors at 550°C, 1 reactor 350°C: 550°C=27cm 350°C=41cm ±0.5°C |
Catalyst Volume (mL)
(isothermal zone)
0.2 – 0.6 mL | 0.4 – 2.0 mL |
0.2 – 0.6 mL | 0.4 – 1.0 mL | 0.4 – 2.0 mL |
0.4 – 1.0 mL |
0.4 – 1.0 mL |
Pressure Ranges (barg)
2 – 80 barg
0.5 – 180 barg (option)
2 – 100 barg
0.5 – 180 barg
2 – 80 barg
0.5 – 180 barg
2 – 20 barg
2 – 50 barg (option)
Reactor Pressure Control
Advanced control RSD ±0.1 barg at reference conditions (gas phase only and 20 barg). For trickle flow Advanced control RSD ±0.5barg.
Standard (±0.5 barg)
Advanced (±0.1 barg) (option)
Standard (±0.5 barg)
Advanced (±0.1 barg) (option)
Advanced (±0.1 barg)
Advanced (±0.1 barg)
Gas Feed Lines
(#Gas Feeds)
Up to 6 + Diluent gas
He, Ar, N2, H2, CH4, CO2, C2H4, C2H6, O2/Inert (≤5%), CO, Other gases
Up to 7 + Diluent gas
He, Ar, N2, H2, CH4, CO2, C2H4, C2H6, O2/Inert (≤5%), CO, Other gases
Up to 7 + Diluent gas
He, Ar, N2, H2, CH4, CO2, C2H4, C2H6, O2/Inert (≤5%), CO, Other gases
Up to 6 + Diluent gas
He, Ar, N2, H2, CH4, CO2, C2H4, C2H6, O2/Inert (≤5%), CO, Other gases
Gas Distribution
(with pressure measurement)
Online Analysis
Full integration GC, MS , GC/MS with data visualisation (option)
Full integration GC, MS , GC/MS with data visualisation
Full integration GC, MS , GC/MS with data visualisation
Full integration GC, MS , GC/MS with data visualisation
Liquid Feed
Split feeding 8 + 8 reators (option)
Pump-Coriolis dosing system
(ambient, cooled)
Pump-Coriolis dosing system
(ambient, cooled, heated 80°C)
Pump-Coriolis dosing system
(ambient, cooled, heated 80°C)
Pump-Coriolis dosing system
(ambient, cooled, heated 80°C)
Liquid Distribution
Microfluidic Distribution
(4-channel glass-chip)
Microfluidics Distribution
(4×4-channel glass-chip)
(16-channel glass-chip)
Active Liquid Distribution (option)
(with automatic isolation valves)
Active Liquid Distribution
(with automatic isolation valves)
Microfluidic Distribution
(4-channel glass-chip)
Liquid Sampling
(G/L Separation)
Parallel liquid sampling (4 x 20ml vials) with sequential on-line gas phase sampling (option)
Automated liquid sampling (4 rows x 16 vials x 8ml) with sequential on-line gas phase sampling (option)
Automated liquid sampling (4 rows x 16 vials x 8ml) with sequential on-line gas phase sampling (option)
Parallel liquid sampling (4 x 20ml vials) with sequential on-line gas phase sampling (option)
Reactors Effluent Handling
(Off-line Analysis Connection)
Full heated circuit up to 180°C with sequential on-line full gas phase sampling (option)
Full heated circuit up to 200°C with sequential on-line full gas phase sampling
Full heated circuit up to 200°C with sequential on-line full gas phase sampling
Full heated circuit up to 200°C with sequential on-line full gas phase sampling
Offline Analysis
Integrated Workflow: SimDist, total S/N, liquid density, balance, label printer, barcode (option)
Integrated Workflow: SimDist, total S/N, liquid density, balance, label printer, barcode
Integrated Workflow: SimDist, total S/N, liquid density, balance, label printer, barcode
Integrated Workflow: SimDist, total S/N, liquid density, balance, label printer, barcode
Waste Handling
Ambient temperature
Heated wax trapping (option)
Ambient temperature / Cooled containers / Heated compartment (wax trapping, heavies)
Ambient temperature / Cooled containers / Heated compartment (wax trapping, heavies)
Ambient temperature / Cooled containers / Heated compartment (wax trapping, heavies)
Safety
Gas sensors and control box (CO, LEL, VOC)
Gas sensors and control box (CO, LEL, VOC)
Gas sensors and control box (CO, LEL, VOC)
Gas sensors and control box (CO, LEL, VOC)
Flowrence® Software
Flowrence® recipe builder, control & database builder
Flowrence® recipe builder, control & database builder
Flowrence® recipe builder, control & database builder